![CNSystems CNAP Monitor 500 HD Operator'S Manual Download Page 120](http://html1.mh-extra.com/html/cnsystems/cnap-monitor-500-hd/cnap-monitor-500-hd_operators-manual_2628121120.webp)
Appendix C
CNAP
®
Monitor 500
Page 17-9
Version 5.2.5,
h
21-FHZU-10002
Recommended separation distance between portable and mobile RF-communication devices and
the CNAP
®
Monitor 500
The CNAP
®
Monitor 500 is intended for use in an electromagnetic environment with controlled RF distur-
bances. The customer or operator of the CNAP
®
Monitor 500 can avoid electromagnetic disturbances by
complying with the minimum distance between portable or mobile RF-communication equipment (transmit-
ter) and CNAP
®
Monitor 500, depending on the power output of the communication equipment as specified
below.
Rated power output
of the transmitter
W
Separation distance depending on the transmitting frequency
m
150 kHz to 80 MHz
P
V
d
∗
=
1
5
,
3
80 MHz to 800 MHz
P
E
d
∗
=
1
5
,
3
800 MHz to 2.5
GHz
P
E
d
∗
=
1
7
0.01
0.12
0.12
0.23
0.1
0.37
0.37
0.74
1
1.17
1.17
2.33
10
3.69
3.69
7.38
100
11.67
11.67
23.33
For transmitters whose maximum rated power output values are not listed in the above list, the minimum
distance can be calculated depending on the transmitting frequency and rated power output by means of
the respective formula, whereas the maximum rated power output is P in watts (W) according to the specifi-
cation of the manufacturer.
Note 1
At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2
These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures,
objects and people.
17.4
Recurrent inspections
The time intervals for the recurrent inspections of the CNAP
®
Monitor 500 as well as the extent of
the work is defined as follows:
•
Every 12 months: Safety control (SC)
•
Every 24 months: Metrological control (MC)
•
Every 24 months: Maintenance (Servicing)
17.4.1
Safety control (SC)
The SC is performed according to IEC 62353 and includes the following activities:
•
Visual inspection of the device
o
damages,
o
hygienic condition and
o
safety relevant labeling (labels, signs)
•
Mechanical inspection (enclosure, handholds, mounts)
•
Functional technical inspection and check
•
Overall assessment
•
Safety test including leakage current measurement acc. EN 60601-1
•
Calibration of NBP module
•
Documentation including: safety protocol, inspection sticker, shipping documents
Distributed by BIOPAC Systems, Inc.