Monitor configuration
CNAP
®
Monitor 500
Page 4-5
Version 5.2.5,
h
21-FHZU-10002
4.2.5
Optimizing Signal Quality
The CNAP
®
Monitor optimizes the signal quality automatically during a measurement. These opti-
mization phases have a duration of several seconds and may lead to atypical wave forms in the
signal view. The optimization phase is marked with the label “Optimizing Signal Quality” in the
bottom field of the CNAP parameter view.
To avoid a potential influence of the signal optimization on a statistical analysis of the beat values,
this mechanism can be deactivated during measurement. This can be done in the menu
Setup |
Service | Smart Mode
– a valid license key needs to be entered in
Service | Monitor | Device
Features
to enable this feature.
To achieve optimal measurement results it is recommended to set
Setup | Service | Smart Mode
| Activation
to
On
– this enables the signal optimization during the measurement (the default
value is
On
).
4.3
Service settings
NOTE:
•
The
Service
menu is divided into 2 layers which can be accessed by entering a
password.
•
You will find the password for the user menu in the CNAP
®
Monitor 500 "Service
manual for users".
•
You will find the password for the service menu in the CNAP
®
Monitor 500 "Instruc-
tions for service" (service manual).
Menu item
Description
Settings
Restore Factory Set-
tings
Restore factory settings
Yes, No
Alarm Defaults…
Enables to adjust alarm limits, reminder, pause and
volume for the patient categories (
Adult, Pediatric
)
within the limits of factory settings. The alterations will
be used for each new measurement. The operator/user
can also restore factory settings.
Submenu
Log…
Lists technical alarms by means of language-
independent error codes
Submenu
Function Tests …
Function tests of the modules IBP analog output, prin-
ter and CNAP/NBP
Submenu
Advanced…
Menu for software update
Submenu
4.4
Feature activation
Starting with software version 3.6 the basic features of the CNAP
®
Monitor 500 can be extended
with a license key. The following features can be activated on the monitor:
•
Additional AUX analog out port (the monitor type is changed to 500at) – v3.7
•
Data recording on USB device – v3.7
•
CNAP
®
blood pressure measurement – v5.2
•
CNAP
®
-PPV measurement(the monitor type is changed to 500i/at+PPV) – v3.6
•
CNAP
®
hemodynamic measurement (the monitor type is changed to 500i/at+HD) –
v5.0