Monitor configuration
CNAP
®
Monitor 500
Page 4-3
Version 5.2.5,
h
21-FHZU-10002
Illustration 19: Trending calibration of blood pressure signal
4.2.2
External calibration
In external calibration mode the CNAP
®
Monitor expects a manually entered blood pressure value
(or height offset – see below) for the calibration of the CNAP
®
signal. After the measurement has
started and the sufficient patient data was entered the menu BP Calibration will be displayed auto-
matically. In this menu the systolic and diastolic blood pressure values can be entered.
Illustration 20: Manual input of blood pressure calibration values
The calibration values can be updated anytime during the measurement in the menu
Parameters
| BP Calibration
.
CAUTION:
•
Make sure that the external NBP reference value has been measured on the patient
immediately before entering it in the menu on the CNAP
®
Monitor. Entering obso-
lete external NBP values results in scaling to a possibly false reference and in show-
ing wrong absolute values!