21-FHZU-10002
Atmospheric pressure
no influence on accuracy of measurement
Measuring time
max. 130 s (ADULT)
Max. inflation time
75 s
Max. cuff pressure
300 mmHg
Automatic deflation after
180 s
Protection against electric
shock
Defibrillation protected applied part – Type BF
Defibrillation protection
According to IEC 60601-1
HF surgery
No restrictions
17.1.4
Printer
Printer
Type
Integrated thermal paper printer
Width
58mm
Roll diameter
60mm
17.2
Connections
BP Wave Out
Bridge supply voltage from
other monitor to CNAP
2 - 10 VDC
Input current max @10V
1.3mA
Sensitivity
5 µV/V/mmHg
V
out
bridge excitation voltage
from CNAP
®
to other monitor
V
in
*5*10
-6
*Pressure [mmHg]