Monitor configuration
CNAP
®
Monitor 500
Page 4-12
Version 5.2.5,
h
21-FHZU-10002
4.7
Output Options
Menu settings for AUX Analog Out and print output:
Menu item
Description
Settings
Analog Out Reference
The value for the reference voltage range oft he AUX
Analog Out interface can be set here.
Default: 0V to 5V
0V to 5V,
-5V to 5V
Analog Out 1
Chanel 1 shows always the BP wave form
Label
Analog Out 2
Select MAP or Pulse for Chanel 2
Default: MAP
MAP,
Pulse
Analog Out 3
Select CO, SV or SVR for output on Chanel 3
Default: CO
CO, SV,
SVR
Analog Out 4
Select Pulse, PPV or SVV for output on Chanel 4
Default: Puls
Pulse,
PPV, SVV
Print On Alarm
Activation of
Print On Alarm
feature
Default: Off
On, Off
Snapshot Delay
Setting of delay time of print reports for
Snapshot
and
Print On Alarm
Default: 10 sec
5 sec, 10
sec, 15
sec
Distributed by BIOPAC Systems, Inc.