Technical alarms and status messages
CNAP
®
Monitor 500
Page 14-9
Version 5.2.5,
h
21-FHZU-10002
14.5
NBP
14.5.1
Status messages
Message
Possible cause
Measures
NBP: Terminated
-
User has stopped cur-
rent NBP measure-
ment
-
n.a.
NBP: Fault
-
Checksum error oc-
curred
-
Start new NBP measurement
-
In case of persistent error, con-
tact service
NBP: Single Measurement
-
User has triggered a
single NBP measure-
ment
-
n.a.
NBP: Automatic Mea-
surement
-
Timed NBP measure-
ment (NBP: Interval)
-
n.a.
NBP:
Checking CNAP
-
NBP check measure-
ment as CNAP
®
blood
pressure changed
more than 25mmHg
within one minute
(compared with last
NBP measurement)
-
n.a.
NBP: Venous Stasis
-
Venous stasis is per-
formed
-
n.a.
14.5.2
Technical alarms
Message
Priority
Possible cause
Measures
NBP:
Weak Or No Signal
Low *
-
Weak or no oscillo-
metric signal
-
NBP measurement
values out of range
-
Check position and fit of NBP
cuff
-
Make sure cuff is placed directly
on the skin
NBP: Artifact
Low *
-
Artifact/irregular
oscillometric signal
-
Check position and fit of NBP
cuff
-
Avoid artifacts (e.g. movement)
-
Check for proper NBP cuff size
-
Check ECG for sinus rhythm