![CNSystems CNAP Monitor 500 HD Operator'S Manual Download Page 57](http://html1.mh-extra.com/html/cnsystems/cnap-monitor-500-hd/cnap-monitor-500-hd_operators-manual_2628121057.webp)
Alarm system
CNAP
®
Monitor 500
Page 6-10
Version 5.2.5,
h
21-FHZU-10002
6.7
Physiological alarms
Alarm message
Priority
Source
Description
Alarm signals
NBP: Sys High
NBP: Dia High
Medium**
NBP
Measured NBP pressure value
exceeds upper alarm limit. In
addition, "Sys", "Dia" indicates
which parameter has exceeded
the alarm limit.
Flashing NBP
values, alarm
message and
acoustic alarm
signal
NBP: Sys Low
NBP: Dia Low
Medium**
NBP
Measured NBP pressure value
falls below lower alarm limit. In
addition, "Sys", "Dia" indicates
which parameter has dropped
below the alarm limit.
Flashing NBP
values, alarm
message and
acoustic alarm
signal
CNAP: Sys High
CNAP: Dia High
CNAP: MAP High
Medium**
CNAP
®
Measured CNAP
®
pressure value
exceeds upper alarm limit. In
addition, "Sys", "MAP", "Dia"
indicates which parameter has
exceeded the alarm limit.
Flashing CNAP
®
values, alarm
message and
acoustic alarm
signal
CNAP: Sys Low
CNAP: Dia Low
CNAP: MAP Low
Medium**
CNAP
®
Measured CNAP
®
pressure value
falls below lower alarm limit. In
addition, "Sys", "MAP", "Dia"
indicates which parameter has
fallen below the alarm limit.
Flashing CNAP
®
values, alarm
message and
acoustic alarm
signal
CNAP: Pulse High
Medium**
CNAP
®
Pulse rate (CNAP
®
) exceeds
upper alarm limit.
Flashing CNAP
®
values, alarm
message and
acoustic alarm
signal
CNAP: Pulse Low
Medium**
CNAP
®
Pulse rate (CNAP
®
) falls below
lower alarm limit.
Flashing CNAP
®
values, alarm
message and
acoustic alarm
signal
NOTE:
•
All technical malfunction alarm messages of the CNAP
®
Monitor 500 or its components
can be found directly in the chapters describing the respective system components.
Distributed by BIOPAC Systems, Inc.