Hemodynamics
CNAP
®
Monitor 500
Page 12-2
Version 5.2.5,
h
21-FHZU-10002
Illustration 55: Activation of the hemodynamic measurement
The permanent activation of the hemodynamic measurement can be set in the setup menu
Setup
| Measurement | Hemodynamic
.
With the activated hemodynamic option the hemodynamic frame on the display gets activated and
the parameters are displayed in the trend display – depending on the settings.
When starting a new CNAP
®
measurement the patient dialog will be displayed automatically. For
calibrating the hemodynamic parameters it is necessary to enter the correct values for gender,
age, weight and height. The patient dialog can be closed by clicking “Ok”.
Illustration 56: Patient dialog shown automatically at the start of a measure-
ment
Approximately 20 seconds after the calibration of the CNAP
®
signal is finished the values for the
hemodynamic parameters are displayed automatically.
Note:
•
Only if all necessary patient data have been entered in the patient dialog (gender,
age, weight and height) the hemodynamic parameters are displayed.
Distributed by BIOPAC Systems, Inc.
Manufactured by CNSystems AG