Appendix C
CNAP
®
Monitor 500
Page 17-5
Version 5.2.5,
h
21-FHZU-10002
AUX (analog output port)
Channel 1
Channel 2
Channel 3
Channel 4
Parameter
BP waveform
MAP
Pulse
CO
SV
SVR
Pulse
PPV
SVV
Range
0 – 500 mmHg
0 – 300 mmHg
0 – 200 bpm
1 – 30 l/min
0 – 200 ml
0 – 5000 dyn s/cm
5
0 – 40 %
0 – 40 %
0 – 200 bpm
Reference
Vol-
tage
configurable: 0 – +5 V or -5 - +5 V
Sampling
fre-
quency
100 Hz
Output Offset
+/- 50 mV
Output Accuracy
5%
Output Internal
Resistor
100 Ohm
Output Current
max. 2 mA
Resolution Im-
pedance
12bit
Isolation (1sec)
4 KVDC min.
Isolation
(>1sec)
1.5 KVDC min.
NOTE:
•
The AUX analog out connector is activated only for the device type CNAP
®
Monitor
500 at.
•
Read the instructions on the info sheet of the AUX analog out connector how to build
up the electrical connection.
Distributed by BIOPAC Systems, Inc.