Monitor configuration
CNAP
®
Monitor 500
Page 4-10
Version 5.2.5,
h
21-FHZU-10002
It is possible ouptut four parameters at a time. The configuration oft he output setting is done in
the menu
Setup | Measurement | Output Options
.
The value for the reference voltage can be selected from the menu
Setup| Measurement | Out-
put Options | Analog Out Reference.
Available reference values are
0V to 5V
and
-5V bis 5V
.
4.6.2
Ethernet
The Ethernet port is restricted for service purposes only.
4.6.3
USB
In addition to service functions such as software updates, the USB port serves for data recording
on a USB flash drive, provided that
Record
for data recording has been activated in
Se-
tup
/Monitor
(
CSV File
or
Advanced
). The settings for
Record
cannot be changed during a
measurement.
NOTE:
•
Do not remove a connected USB flash drive during an ongoing measurement.
•
Only USB flash drives formatted with FAT32 are supported by the CNAP
®
Monitor
500 for data storage and software updates.
•
Recording data as CSV files is an optional feature. If the entry
CSV File
in the
menu
Setup | Monitor | Record
is not visible this feature is not available on
your device. How to activate optional features read the description in chapter 4.4.
CAUTION:
•
Do not connect any input devices or devices with high power consumption to the
USB port. Only USB flash drives are supported.
STOP:
•
Because of a possible influence on the patient safety it is not allowed to connect USB
hard drives or any other devices using external power supplies to the CNAP
®
Monitor
500.