Technical alarms and status messages
CNAP
®
Monitor 500
Page 14-11
Version 5.2.5,
h
21-FHZU-10002
Message
Priority
Possible cause
Measures
-
In case of persistent error, con-
tact service
NBP: ADC Failure
Low *
-
Analog/digital con-
verter out of mea-
suring range
-
Reboot CNAP
®
Monitor 500
-
In case of persistent error, con-
tact service
NBP: Calibration
Failure
Low *
-
Faulty EEPROM
calibration data
-
Reboot CNAP
®
Monitor 500
-
In case of persistent error, con-
tact service
Distributed by BIOPAC Systems, Inc.
Manufactured by CNSystems AG