144
Maintenance manual
2016/02
AlveoPC
Before opening the Mixer, you must switch off and disconnect
its power supply.
1.
Remove the Mixer casing.
2.
Reconnect the main power cord.
3.
Using the extension cable provided, connect the cover safety
device cable.
4.
Turn on the instrument's switch.
230VAC present inside the instrument.
Use a voltmeter and the diagram on the next page below to check
the voltages on the Mixer CPU board.
5.
Switch off the instrument and disconnect it.
6.
Put the casing back on.
B13: Adjustments and checks for the Mixer
Operation B13-3 Electrical checks on the Alveo CPU board
Page: 1/2
Consumables:
Tools:
-
Voltmeter
Extension cable for cover safety device (see Chopin)
Summary of Contents for AlveoPC
Page 2: ......
Page 4: ...AlveoPC 4 Maintenance manual...
Page 6: ...AlveoPC 6 Maintenance manual...
Page 8: ...8 Maintenance manual 2016 02 AlveoPC...
Page 17: ...17 Maintenance manual 2016 02 AlveoPC Chapter 2 TROUBLE SHOOTING...
Page 18: ...18 Maintenance manual 2016 02 AlveoPC...
Page 58: ...58 Chapter 2 Trouble shooting Maintenance manual 2016 02 AlveoPC...
Page 59: ...59 Maintenance manual 2016 02 AlveoPC Chapter 3 REPAIRS...
Page 60: ...60 Maintenance manual 2016 02 AlveoPC...
Page 66: ...66 Chapter 3 Repairs Maintenance manual 2016 02 AlveoPC 2 Instructions See following pages...
Page 67: ...Chapter 3 Repairs 67 Maintenance manual 2016 02 AlveoPC A Alveograph...
Page 111: ...111 Maintenance manual 2016 02 AlveoPC B Mixer...
Page 112: ...112 Maintenance manual 2016 02 AlveoPC...
Page 146: ...146 Maintenance manual 2016 02 AlveoPC...
Page 147: ...147 Maintenance manual 2016 02 AlveoPC Chapter 4 SPARE PARTS...
Page 148: ...148 Maintenance manual 2016 02 AlveoPC...
Page 150: ...150 Chapter 4 Spare parts Maintenance manual 2016 02 AlveoPC...
Page 152: ...152 Chapter 4 Spare parts Maintenance manual 2016 02 AlveoPC figure 1 Alveograph 1 2 5 3 4...
Page 174: ...174 Chapter 4 Spare parts Maintenance manual 2016 02 AlveoPC...