- 79 -
8-17. Compressed/Expanded Printing Settings “Cmp&Exp”
When “Comp. &Exp.Print” is selected for recording format, set the related items. A specified area within the chart
recording range can be shrunk or expanded.
• Individual setting available for channels.
• Recording range can be set arbitrarily regardless of the setting of range/chart recording upper and lower limits.
• Up to two break points can be set, therefore three shrunk or expanded areas can be obtained at maximum.
▼
indicates a set point.
Example of compressed/expanded printing
settings
Entire
recording
range
Scale
(1st break point)
(2nd break point)
Compressed
Expanded
Compressed
Recording
position
-200
1370
-100 600
800
1200
0%
100%
20% 80%
-100 600
800 1200
For the case of K thermocouple
Measuring range