69
Senior Solutions™ Therapy System
6 REMOVAL/REPLACEMENT
6.12 CHANNEL 3/4 ELECTROTHERAPY MODULE
CONNECTOR
BOARD
A. Tools and Equipment Required
#1 Phillips Screwdriver
Flat Blade Screwdriver
B. Module Connector Board Removal
Remove Module from Therapy System.
Refer to 6.2, part C.
Release the Module Top Retaining Tabs with
the Flat Blade Screwdriver.
See Figure 6.45
Remove Top.
Remove Module Connector Infill
. See
Figure 6.46.
Carefully pull Connector Board out toward
front of Module.
See Figure 6.47.
C. Replacing Connector Board
Replace new Connector Board, part
number 27059 in reverse order of steps
1-4 above. Make Certain the words
“THIS SIDE UP” on Connector Board are
facing up. Make certain Connector Board is
completely seated in Stim Board Connector.
See Figure 6.47.
Re-install the Connector Infill and Lanyard.
Refer to Figure 6.46.
Re-assemble Therapy System referring to
the appropriate sections of this manual for
proper instructions.
NOTE:
Do not over-tighten the screws. Over-
tightening will damage the threads of the
brass inserts.
1.
2.
1.
2.
3.
4.
5.
1.
2.
3.
FIGURE 6.45
FIGURE 6.46
RELEASE TABS
LIFT CONNECTOR INFILL OUT
OF MODULE HOUSING
FIGURE 6.47
DISCONNECT THE SYSTEM FROM THE POWER
SOURCE (OUTLET OR REMOVE BATTERY MODULE IF
INSTALLED) BEFORE ATTEMPTING ANY MAINTENANCE,
INSTALLATION, REMOVAL, OR REPLACEMENT
PROCEDURES TO PREVENT ELECTRICAL SHOCK AND
POSSIBLE DAMAGE TO SYSTEM.
Summary of Contents for SENIOR SOLUTIONS
Page 1: ...ISO 13485 CERTIFIED Therapy System SERVICEMANUAL Moving Rehabilitation Foward ...
Page 2: ......
Page 5: ......
Page 105: ......
Page 106: ......