Senior Solutions™ Therapy System
5 TROUBLESHOOTING
5.13 MICROCURRENT PROBE MODE TEST
(CONTINUED)
B. MicroCurrent Probe Mode Test Results
Unit performs as described in
steps 5-7
and 10.
Unit passed test.
No Search Mode beep. Unit failed test. Try a
known good Microcurrent Probe and repeat
test. If still no beep. Replace appropriate
Stim Board.
Good Search Mode beep but no beep
speed increase when probe touched to
metal tip of the Black (-) Leadwire.
Unit failed test.
Replace probe, and lead wire with known
good probe, and lead wire. Repeat test and
use process of elimination to determine if
probe,or lead wire is the problem. Same
results after test is repeated, replace
appropriate Stim Board.
Tones OK but no green or gray vertical bar
in Contact Quality Scale. Unit failed test.
Replace appropriate Stim Board. If problem
persists, replace Control Board.