60
Senior Solutions™ Therapy System
6 REMOVAL/REPLACEMENT
FIGURE 6.21
REMOVE
HARNESSES
Lay System Top on the edge and disconnect
the remaining harnesses from the Control
Board.
See Figure 6.21.
C. Replacing Top to Bottom
Replace System Top by reversing the
steps 1- 5
on page 59.
NOTE:
Do not overtighten the screws. Overtightening
will damage the threads of the brass inserts.
6.
DISCONNECT THE SYSTEM FROM THE POWER
SOURCE (OUTLET OR REMOVE BATTERY MODULE IF
INSTALLED) BEFORE ATTEMPTING ANY MAINTENANCE,
INSTALLATION, REMOVAL, OR REPLACEMENT
PROCEDURES TO PREVENT ELECTRICAL SHOCK AND
POSSIBLE DAMAGE TO SYSTEM.