Senior Solutions™ Therapy System
6 REMOVAL/REPLACEMENT
FIGURE 6.42
FIGURE 6.43
100 WATT
SCREWS
FIGURE 6.44
MAINS WIRING
HARNESS
PLIERS NOT
SHOWN
FOR CLARITY
75 WATT
SCREWS
DISCONNECT THE SYSTEM FROM THE POWER
SOURCE (OUTLET OR REMOVE BATTERY MODULE IF
INSTALLED) BEFORE ATTEMPTING ANY MAINTENANCE,
INSTALLATION, REMOVAL, OR REPLACEMENT
PROCEDURES TO PREVENT ELECTRICAL SHOCK AND
POSSIBLE DAMAGE TO SYSTEM.
Using Insulated Needle Nose Pliers,
disconnect the Mains Connector Wiring
Harness from the Power Supply Assembly.
See Figure 6.42.
Remove 100 Watt Power Supply from
Mounting Bracket by removing the two
retaining screws from the inside top of the
Power Supply.
See Figure 6.43.
NOTE:
Steps 10 and 11
below apply only to
Combination Systems.
Remove 75 Watt Power Supply from
Mounting Bracket by removing the two
mounting screws on the back of Mounting
Bracket securing the 75 Watt Power Supply.
See inset at
Figure 6.43.
Using Insulated Needle Nose Pliers, remove
the 100 Watt Power Supply Harness from
the 75 Watt Power Supply.
See Figure
6.44.
C. Replacing Power Supplies
Replace new Power Supplies in reverse
order of preceding steps using part
numbers:
a)
27048 100 Watt Power Supply
b)
27049 75 Watt Power Supply (Combo
System
Only)
Re-assemble Therapy System.
Refer to
6.10, part C, 6.9, part C, 6.8, part C,
and 6.4, part C
for proper instructions.
NOTE:
Do not over-tighten the screws. Over-
tightening will damage the threads of the
brass inserts.
8.
9.
10.
11.
1.
2.