Senior Solutions™ Therapy System
5 TROUBLESHOOTING
CRITICAL ERRORS DEMANDING TECHNICAL SERVICE 300399
5.1 THERAPY SYSTEM ERROR MESSAGES (CONTINUED)
Code
Number
Type
Message
Probable Cause
Possible Remedies
308
Critical Error
Error reading from Stim Board.
A.
On Therapy System, make certain internal Ribbon Cable is seated on Stim PC Board and Control
Board.
B.
On Channel 3/4 Electrotherapy Module, make certain the Therapy System to Module Ribbon
Cable is seated in the Therapy System and on the Module.
C.
If problem persists, replace appropriate Stim PC Board.
D.
Replace Control Board.
309
Critical Error
Error reading from Stim Board.
A.
On Therapy System, make certain internal Ribbon Cable is seated on Stim PC Board and Control
Board.
B.
On Channel 3/4 Electrotherapy Module, make certain the Therapy System to Module Ribbon
Cable is seated in the Therapy System and on the Module.
C.
If problem persists, replace appropriate Stim PC Board.
D.
Replace Control Board.
310
Critical Error
Ultrasound Board has reported an error.
A.
Make certain Ultrasound PC Board is completely seated on internal Header and Ribbon Cable is
seated on Stim PC Board and Control Board.
B.
If problem persists, replace Ultrasound PC Board.
C.
Replace Control Board.
311
Critical Error
Ultrasound Board has reset.
A.
Make certain Ultrasound PC Board is completely seated on internal Header and Ribbon Cable is
seated on Stim PC Board and Control Board.
B.
If problem persists, replace Ultrasound PC Board.
C.
Replace Control Board.
312
Critical Error
Error writing to Ultrasound Board.
A.
Make certain Ultrasound PC Board is completely seated on internal Header and Ribbon Cable is
seated on Stim PC Board and Control Board.
B.
If problem persists, replace Ultrasound PC Board.
C.
Replace Control Board.
313
Critical Error
Error reading from Ultrasound Board.
A.
Make certain Ultrasound PC Board is completely seated on internal Header and Ribbon Cable is
seated on Stim PC Board and Control Board.
B.
If problem persists, replace Ultrasound PC Board.
C.
Replace Control Board.
314
Critical Error
Error reading from Ultrasound Board.
A.
Make certain Ultrasound PC Board is completely seated on internal Header and Ribbon Cable is
seated on Stim PC Board and Control Board.
B.
If problem persists, replace Ultrasound PC Board.
C.
Replace Control Board.
315
Critical Error
Error reading from Ultrasound Board.
A.
Make certain Ultrasound PC Board is completely seated on internal Header and Ribbon Cable is
seated on Stim PC Board and Control Board.
B.
If problem persists, replace Ultrasound PC Board.
C.
Replace Control Board.
316
Critical Error
Error writing to Module Board.
A.
Make certain the Therapy System to Module Ribbon Cable is seated in the Therapy System and
on the Module.
B.
If problem persists, replace appropriate Stim PC Board.
C.
Replace Control Board.
317
Critical Error
Error reading from Module Board.
A.
Make certain the Therapy System to Module Ribbon Cable is seated in the Therapy System and
on the Module.
B.
If problem persists, replace appropriate Stim PC Board.
C.
Replace Control Board.
318
Critical Error
Error reading from Module Board.
A.
Make certain the Therapy System to Module Ribbon Cable is seated in the Therapy System and
on the Module.
B.
If problem persists, replace appropriate Stim PC Board.
C.
Replace Control Board.