6 REMOVAL/REPLACEMENT
FIGURE 6.7
4 MM X 20 MM
SCREWS
FIGURE 6.8
EXTENDED FRONT
ACCESS PANEL
LANYARD
FIGURE 6.9
DISCONNECT THE SYSTEM FROM THE POWER
SOURCE (OUTLET OR REMOVE BATTERY MODULE IF
INSTALLED) BEFORE ATTEMPTING ANY MAINTENANCE,
INSTALLATION, REMOVAL, OR REPLACEMENT
PROCEDURES TO PREVENT ELECTRICAL SHOCK AND
POSSIBLE DAMAGE TO SYSTEM.
Holding Module to system, position both
on one side and secure the Module to the
System with four 4 mm x 20 mm Screws.
See Figure 6.7.
Tighten only enough to
prevent Module from shifting on Therapy
System.
Set the assembly upright on the work
surface and install the new Extended Front
Access Panel to the Lanyard.
See Figure
6.8.
NOTE:
When mounting the Front Access Panel
to the Therapy System, make certain the
Lanyard does not become kinked.
Route Power Cord through Module to
Therapy System.
See Figure 6.9.
Re-install Rear Panel. Plug Power Cord into
an approved power outlet.
NOTE:
If installing Therapy System with Module to
a Therapy System Cart, refer to the Therapy
System Cart installation instructions.