Senior Solutions™ Therapy System
6 REMOVAL/REPLACEMENT
Turn the System On using the On/Off Switch.
The system will automatically recognize the
added module and display a configuration
change message.
See Figure 6.16.
Read and carefully follow the instructions on
the screen.
C. Module Removal
Remove Module, if necessary, from Therapy
System.
Refer to 6.2, part C
for instructions.
Place a Flat Blade Screwdriver under the
locking tab of the sEMG Module. Firmly push
in and twist Screwdriver. The Module should
release from the Therapy System.
See Figure
6.17.
NOTE:
If no sEMG Module will be re-installed on the
system install sEMG Plug Kit, part number
28027, to protect and cover the Stim Board
contacts. The Therapy System may be placed
back into service without an sEMG Module
installed as long as the sEMG Plug Kit is
installed to protect the PC Board contacts.
Refer to
steps 7 and 8
to finalize removal.
7.
8.
1.
2.
3.
FIGURE 6.16
FIGURE 6.17
Verify that the module installed is the module
displayed in the message BEFORE pressing the
START Button. If it is not, DO NOT press the START
Button. Turn the system OFF and back ON. If the
problem persists, call the dealer or Chattanooga
Group Technical Support immediately. DO NOT USE
THE SYSTEM until all necessary repairs are made by
a Technician certified by Chattanooga Group. If use
is attempted before repairs are made, the system
may operate unpredictably and has the potential
of causing injury to the patient or damage to the
system internal components.