Senior Solutions™ Therapy System
6 REMOVAL/REPLACEMENT
6.8 THERAPY SYSTEM CONNECTOR BOARD
A. Tools and Equipment Required
#1 Phillips Screwdriver
B. Connector Board Removal
Separate Top from Bottom.
Refer to 6.4,
part B.
Remove Connector Infill and Lanyard. Lay
aside.
See Figure 6.31.
Carefully pull Connector Board out toward
front of System.
See Figure 6.32.
C. Replacing Connector Board
Replace new Connector Board, part number
27059, in reverse order of
steps 1-3
above.
Make Certain the words “THIS SIDE UP” on
Connector Board are facing up. Make certain
Connector Board is completely seated in Stim
Board Connector.
See Figure 6.33.
Re-install the Connector Infill and Lanyard.
Refer to Figure 6.31.
Re-assemble Therapy System.
Refer to 6.4,
part C
for proper instructions.
NOTE:
Do not over-tighten the screws. Over-
tightening will damage the threads of the
brass inserts.
1.
2.
3.
1.
2.
3.
FIGURE 6.31
FIGURE 6.32
REMOVE CONNECTOR INFILL
LANYARD NOT ILLUSTRATED
PULL CONNECTOR BOARD
TOWARD SYSTEM FRONT
FIGURE 6.33
DISCONNECT THE SYSTEM FROM THE POWER
SOURCE (OUTLET OR REMOVE BATTERY MODULE IF
INSTALLED) BEFORE ATTEMPTING ANY MAINTENANCE,
INSTALLATION, REMOVAL, OR REPLACEMENT
PROCEDURES TO PREVENT ELECTRICAL SHOCK AND
POSSIBLE DAMAGE TO SYSTEM.