71
Senior Solutions™ Therapy System
6 REMOVAL/REPLACEMENT
6.14 MOUNTING AND DISMOUNTING THERAPY
SYSTEM AND THERAPY SYSTEM CART
A. Mounting Therapy System to
Therapy System Cart
Remove all the Storage Bins from both
sides of the Therapy System Cart by pulling
each bin out and up.
See Figure 6.49.
Allow approximately 4.5in (11.5 cm) of the
power cord extending through the top of
the cart for connecting to system. If the
system is equipped with an optional NiMH
Battery, or Channel 3/4 Electrotherapy
Module, it will be necessary to allow 6.5
in (16.5 cm) of the Power Cord extending
through the top of the Therapy System
Cart.
See Figure 6.50.
Secure the System to the cart with the four
socket head screws in the Therapy System
Cart Top.
See Figure 6.51.
NOTE:
Secure the System to the cart by tightening
the screws by hand only. Do not use
a wrench to tighten the screws. Over-
tightening may cause damage to the
System or Module housing.
Plug Power Cord into the System Mains
Disconnect and reinstall the Rear Access
Panel. Install all lead wires and cables to
the System.
Install Storage Bins into Therapy System
Cart. START with bottom Storage Bin first.
B. Dismounting Therapy System from
Therapy System Cart
To remove the System from the Therapy
System Cart, repeat the Mounting System
to Therapy System Cart instructions in
reverse order.
1.
2.
3.
4.
5.
1.
FIGURE 6.49
FIGURE 6.50
FIGURE 6.51
4.5 IN 11.5 CM
SYSTEM ONLY
6.5 IN 16.5 CM
SYSTEM W/MODULE
REMOVE
ALL BINS
HAND TIGHTEN THE SOCKET
HEAD SCREWS TO SECURE
SYSTEM TO CART
DISCONNECT THE SYSTEM FROM THE POWER
SOURCE (OUTLET OR REMOVE BATTERY MODULE IF
INSTALLED) BEFORE ATTEMPTING ANY MAINTENANCE,
INSTALLATION, REMOVAL, OR REPLACEMENT
PROCEDURES TO PREVENT ELECTRICAL SHOCK AND
POSSIBLE DAMAGE TO SYSTEM.