6 REMOVAL/REPLACEMENT
FIGURE 6.5
BLUE STRIP
RIBBON CABLE MUST BE
POSITIONED AS SHOWN
POSITION THERAPY SYSTEM
AND MODULE AS SHOWN
FIGURE 6.6
FEET ALIGNED
WITH INDENTIONS
RIBBON CABLE MUST BE
POSITIONED AS SHOWN
Lay Therapy System on its side and
position the Module to be installed beside
it. Install Ribbon Cable to Module and
seat completely on module pins. Make
certain the Blue Strip on Ribbon Cable is
positioned as shown in
Figure 6.5.
Position Therapy System over Module
as shown in
Figure 6.6.
Align Therapy
System Feet with Module Indentions.
RIBBON CABLE MUST BE POSITIONED AS
SHOWN
!
Place System onto Module making certain
the System Feet are within the Module
Identions.
7.
8.
9.
DISCONNECT THE SYSTEM FROM THE POWER
SOURCE (OUTLET OR REMOVE BATTERY MODULE IF
INSTALLED) BEFORE ATTEMPTING ANY MAINTENANCE,
INSTALLATION, REMOVAL, OR REPLACEMENT
PROCEDURES TO PREVENT ELECTRICAL SHOCK AND
POSSIBLE DAMAGE TO SYSTEM.
DO NOT TWIST RIBBON CABLE!
If Ribbon Cable is twisted, the pins will not properly
align. If power is applied to the system with
misalignment of pins or a twisted ribbon cable,
the controlling electronics in the Module will be
destroyed and possible damage to the System
internal components could occur.