Spirit™ Bed Owner’s Manual – 1-70-001 – Rev D
© Carroll Hospital Group Inc.
Page 90 of 110
Cable. If pendant does not work in new port, continue to next
step
5.
Test pendant on a properly functioning bed (if available). If
pendant works, replace DJB Box and T-Cable. If pendant doesn’t
work, replace pendant
Only some functions working
on the rail controls or
pendant
1.
Some functions may be locked out. Check patient lock-out
buttons on footboard. Amber LED on patient lock-out button
prevents the bed operation from the rail controls and pendant
2.
Check cable connection from rail control to port on bed frame
3.
Check pendant connection into DJB Box under head section of the
mattress deck
4.
Replace rail control or pendant
Bed will not lower all the way
1.
Check to see if the Bed UP/DOWN function is locked out on the
footboard
2.
Check to see if bed is in Trendelenburg or Reverse Trendelenburg
position. For patient safety, bed remains at the angle established
by staff whenever the bed is raised and lowered. If the bed is in
TREND/rev.TREND, it will not lower past the preset TREND
/rev.TREND safety zone (minimum mattress deck elevation of 15”).
If the bed is angled, level bed by pressing opposite TREND button.
The bed will automatically stop when the mattress deck is in the
horizontal position
HINT: A slight Trendelenburg or Reverse Trendelenburg angle may not
be easily visible. If the bed appears “stuck” at a bed elevation of
approximately 15”, briefly press and hold the TREND button to achieve a
slight bed angle, then hold the rev.TREND button until the bed
automatically stops at level. Then, lower the bed using the Bed DOWN
button
OR
Use the “Hi-Lo Actuator Synchronization Feature”. Lower the bed
by pressing and holding the Bed DOWN button on either the staff rail
control or the pendant. The bed will first level the mattress deck. Once
level, the bed will automatically begin to lower
4.6
Troubleshooting Guide for Unique Features of Spirit Select™ Bed
Problem / Failure
Recommended Action
Footboard staff control does
not power up (no icons
illuminated on Power Status
Indicator and/or display does
not activate)
1.
Check to see if the bed is operable. If not operable, check to
ensure bed is connected to AC power. If not plugged in, connect
to AC power (wall outlet) and allow battery to recharge
2.
Check to see if the bed is operable from either rail controls or
pendant (if equipped). If bed is operable, continue to next step. If
the bed is not operable from rail controls, check rail and pendant
T-cable connections at rail and in CB10 control box
3.
If bed is operable from footboard control, continue to next step.
If the bed is not operable from the footboard control check above
sections. Check footboard staff control cable connection into CB10
control box
4.
If bed remains inoperable, check to see if a supply cable has been
disconnected from the CB09 and/or the CB10 control box.
Summary of Contents for Spirit
Page 2: ...Spirit Bed Owner s Manual 1 70 001 Rev D Carroll Hospital Group Inc Page 2 of 110 ...
Page 10: ...Spirit Bed Owner s Manual 1 70 001 Rev D Carroll Hospital Group Inc Page 10 of 110 ...
Page 22: ...Spirit Bed Owner s Manual 1 70 001 Rev D Carroll Hospital Group Inc Page 22 of 110 ...
Page 82: ...Spirit Bed Owner s Manual 1 70 001 Rev D Carroll Hospital Group Inc Page 82 of 110 ...