Patient Weight Display Units – Units Locked
KG units used for example
Once the patient weight display units have been
selected (ENTER button pushed) you will be prompted to
lock the units.
If you do wish to lock the patient weight display units,
simply use the UP/DOWN Arrow buttons to scroll to the
“YES” option.
1.
Press ENTER button to select the lock units. The
display will automatically go to the “UNIT
LOCKED” screen to show that the unit change
has been locked. The display will then
automatically return to the “LOCK LBS/KG”
screen of the Service Menu.
2.
Use the UP/DOWN Arrow buttons to scroll to the
“EXIT MENU” screen of the Service Menu.
3.
Press ENTER button to exit the Service Menu to
the home screen.
IMPORTANT: If the patient weight display unit is
locked at this point, it cannot be changed from the Main
Menu. Staff will need to access the Service Menu to
change the patient weight display units.