Spirit™ Bed Owner’s Manual – 1-70-001 – Rev D
© Carroll Hospital Group Inc.
Page 76 of 110
3.12.2
Instructions: Scale Calibration
1.
The first step of the scale calibration procedure is to perform the bed angle calibration. See page
80 for instructions.
2.
After the successful completion of the bed angle calibration, return to this page to continue with
scale calibration procedure.
3.
Elevate the bed to a comfortable working height.
Preparing the Bed for Calibration
Immediately following the successful completion of the
bed angle calibration procedure, you will be instructed to
clear the bed of all equipment (mattress, lines, etc).
Ideally, the mattress deck should be bare.
IMPORTANT: Do not place the calibration masses on
the bed at this point.
Set Calibration Units and Weight
KG units used in this example
You will now be asked to set the calibration weight. This
is the total sum of the certified calibration masses used
to the calibration procedure.
1.
Press ENTER button to begin.
2.
Use the UP/DOWN Arrow buttons to select the
units that correspond to the calibration masses
that will be used.
3.
Press ENTER button to select the appropriate
units. You will automatically be taken to the
calibration weight input screen.
4.
Each time the UP/DOWN Arrow buttons are
pushed, an increase/decrease of 1 LB or 1 KG
(depending upon which unit was just selected)
will register.
5.
After the appropriate calibration mass has been
inputted, press ENTER button to set calibration
weight.
IMPORTANT: Correctly setting the calibration weight
and then using that exact total sum of calibration mass
is essential to ensure the accuracy of the scale. For this
reason, Carroll Hospital Group strongly recommends the
use of certified calibration masses.
From page 74
Begin Scale
Calibration
Procedure
AND
AND
Summary of Contents for Spirit
Page 2: ...Spirit Bed Owner s Manual 1 70 001 Rev D Carroll Hospital Group Inc Page 2 of 110 ...
Page 10: ...Spirit Bed Owner s Manual 1 70 001 Rev D Carroll Hospital Group Inc Page 10 of 110 ...
Page 22: ...Spirit Bed Owner s Manual 1 70 001 Rev D Carroll Hospital Group Inc Page 22 of 110 ...
Page 82: ...Spirit Bed Owner s Manual 1 70 001 Rev D Carroll Hospital Group Inc Page 82 of 110 ...