To Operate the Bed:
Use the footboard staff control, rail controls, or the pendant to raise or lower the head
section, the knee section, or the entire bed.
To Operate One-Touch CPR:
Push and release the CPR button; head and knee sections will automatically flatten.
To Operate Trendelenburg:
Push and hold the TREND (Trendelenburg) or rev.TREND (Reverse Trendelenburg)
button on the footboard control to achieve either function. To automatically level the
bed, push and hold the opposite function until the bed stops.
To Operate Chair:
Push and hold the CHAIR button. Green LED illuminates when bed angle is attained,
then head and knee will raise to full extension. Push and hold button again to remove
chair. (Bed will reverse its position each time the CHAIR button is pressed).
To Operate Auto Contour:
Press the CONTOUR button so that the green light is on. Knee section will raise
automatically when head section is raised. (NOTE: knee lock-out will override Auto
Contour).
Lock-Outs:
The side rail controls and pendant can be locked or unlocked (to restrict patient use).
Lock-outs for each bed function are located on the footboard staff control underneath
each function. When the LOCK button LED is illuminated, patient control of that
particular function with the side rail controls or pendant is restricted. If the LOCK
button LED is not illuminated, then patient control of that function is permitted.
Master Lock-Out:
To restrict all side rail, pendant,
and
footboard controls, press all three LOCK buttons
simultaneously (the LOCK button LED’s will sequentially flash). When Master Lock-
Out has been activated, the emergency CPR function and Nurse Call remains enabled.
To Operate Quad Rails:
To lower the rails, gently depress the green push button, then rotate the rail
down. Head rails rotate toward the headboard, hips rails rotate toward the
footboard. The rail will automatically lock in the UP position.
To Move and Lock the Bed:
Use the Central “Lock & Steer” system:
BRAKE: All four wheels are locked in “BRAKE” mode
STEER: Foot-end wheels lock in “STEER” mode, head-end wheels swivel freely
NEUTRAL: All four wheels swivel freely in “NEUTRAL” mode
Summary of Contents for Spirit
Page 2: ...Spirit Bed Owner s Manual 1 70 001 Rev D Carroll Hospital Group Inc Page 2 of 110 ...
Page 10: ...Spirit Bed Owner s Manual 1 70 001 Rev D Carroll Hospital Group Inc Page 10 of 110 ...
Page 22: ...Spirit Bed Owner s Manual 1 70 001 Rev D Carroll Hospital Group Inc Page 22 of 110 ...
Page 82: ...Spirit Bed Owner s Manual 1 70 001 Rev D Carroll Hospital Group Inc Page 82 of 110 ...