Spirit™ Bed Owner’s Manual – 1-70-001 – Rev D
© Carroll Hospital Group Inc.
Page 18 of 110
DANGER
EXPLOSION Hazard
DO NOT operate this bed in the presence of a flammable anaesthetic mixture with air or
with nitrous oxide.
DANGER
Possible FIRE Hazard exists if the handheld pendant is not disabled when using oxygen
administering equipment. If bed is equipped with a pendant, disconnect the pendant
from the port at the side of the bed and insert pendant into its holster or attach pendent
to bed linen using the included linen clip when not in use.
Keep pendant cord clear of moving equipment.
DANGER
Possible FIRE Hazard exists when connecting the electric cord to a 120V wall outlet if
oxygen administering equipment is in use.
Use either a patient nasal mask or ½-length tent, ensuring the tent never extends below
the mattress deck, when administering oxygen.
WARNING
DO NOT roll the bed over the power supply cord. DO NOT entangle the power supply
cord on other objects. A pinched power supply cord can become damaged is could be
dangerous. Be aware of the power supply cord location especially when moving the bed.
WARNING
This equipment may cause electromagnetic disturbances to its environment. The bed
has NOT been evaluated for electromagnetic compatibility (EMC).
The performance of this equipment may be degraded when exposed to electromagnetic
disturbances and/or electrostatic discharge.
Summary of Contents for Spirit
Page 2: ...Spirit Bed Owner s Manual 1 70 001 Rev D Carroll Hospital Group Inc Page 2 of 110 ...
Page 10: ...Spirit Bed Owner s Manual 1 70 001 Rev D Carroll Hospital Group Inc Page 10 of 110 ...
Page 22: ...Spirit Bed Owner s Manual 1 70 001 Rev D Carroll Hospital Group Inc Page 22 of 110 ...
Page 82: ...Spirit Bed Owner s Manual 1 70 001 Rev D Carroll Hospital Group Inc Page 82 of 110 ...