EMC Declaration Tables
70-00082-01 A
Atria Electrocardiograph Operating Instructions
11-13
Recommended Separation Distances
Refer to the following table for recommended separation distances between
the Atria and portable and mobile RF communications equipment.
The Atria is intended for use in an electromagnetic environment in which
radiated RF disturbances are controlled. The user of the Atria can help
prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment (transmitters)
and the Atria as recommended below, according to the maximum output
power of the communications equipment.
Rated maximum output
power of transmitter
W
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
d
= 1.2
√
P
80 MHz to 800 MHz
d
= 1.2
√
P
800 MHz to 2,5 GHz
d
= 2.3
√
P
0,01
0,12
0,12
0,23
0,1
0,38
0,38
0,73
1
1,2
1,2
2,3
10
3,8
3,8
7,3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the
recommended separation distance d in metres (m) can be estimated using the
equation applicable to the frequency of the transmitter, where P is the maximum
output power rating of the transmitter in watts (W) according to the transmitter
manufacturer.
NOTE1: At 80 MHz and 800 MHz, the separation distance for the higher frequency
range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects and
people.
Summary of Contents for Atria 3100
Page 8: ...viii Atria Electrocardiograph Operating Instructions 70 00082 01 A...
Page 22: ...2 Equipment Setup 2 8 Atria Electrocardiograph Operating Instructions 70 00082 01 A...
Page 34: ...3 User Interface 3 12 Atria Electrocardiograph Operating Instructions 70 00082 01 A...
Page 72: ...6 Patient Preparation 6 10 Atria Electrocardiograph Operating Instructions 70 00082 01 A...
Page 82: ...7 Acquiring ECG Reports 7 10 Atria Electrocardiograph Operating Instructions 70 00082 01 A...
Page 94: ...8 Patient Records 8 12 Atria Electrocardiograph Operating Instructions 70 00082 01 A...
Page 104: ...9 Sample Printouts 9 10 Atria Electrocardiograph Operating Instructions 70 00082 01 A...
Page 112: ...10 Troubleshooting 10 8 Atria Electrocardiograph Operating Instructions 70 00082 01 A...
Page 128: ...11 Maintenance and Service 11 16 Atria Electrocardiograph Operating Instructions 70 00082 01 A...
Page 134: ...4 Atria Electrocardiograph Operating Instructions 70 00082 01 A...