Recognizing and Reducing ECG Artifacts
70-00082-01 A
Atria Electrocardiograph Operating Instructions
10-7
To reduce electrical interference:
• Keep the power cord away from the patient and patient cable.
• Connect the Atria to a properly grounded wall outlet.
• Arrange the patient cable leads together, closely following
the body contour.
• Turn on line filter. Check
[Setup]
and then select
Waveform
Preferences
>
Line Filter
. For more information, see “Line
Filter” on pg. 4-8.
• Ensure Diathermy or X-ray equipment in adjacent rooms is
not operating. Other electrical equipment including electric
beds, televisions, and lighting fixtures may also generate
interference (even when not in use).
• Try moving the patient to another place in the room.
Sometimes, electrical wiring in walls and ceilings causes
interference.
• Operate the Atria from battery power.
Summary of Contents for Atria 3100
Page 8: ...viii Atria Electrocardiograph Operating Instructions 70 00082 01 A...
Page 22: ...2 Equipment Setup 2 8 Atria Electrocardiograph Operating Instructions 70 00082 01 A...
Page 34: ...3 User Interface 3 12 Atria Electrocardiograph Operating Instructions 70 00082 01 A...
Page 72: ...6 Patient Preparation 6 10 Atria Electrocardiograph Operating Instructions 70 00082 01 A...
Page 82: ...7 Acquiring ECG Reports 7 10 Atria Electrocardiograph Operating Instructions 70 00082 01 A...
Page 94: ...8 Patient Records 8 12 Atria Electrocardiograph Operating Instructions 70 00082 01 A...
Page 104: ...9 Sample Printouts 9 10 Atria Electrocardiograph Operating Instructions 70 00082 01 A...
Page 112: ...10 Troubleshooting 10 8 Atria Electrocardiograph Operating Instructions 70 00082 01 A...
Page 128: ...11 Maintenance and Service 11 16 Atria Electrocardiograph Operating Instructions 70 00082 01 A...
Page 134: ...4 Atria Electrocardiograph Operating Instructions 70 00082 01 A...