Sensor Application
70-00082-01 A
Atria Electrocardiograph Operating Instructions
6-3
Sensor Application
Apply sensors according to the instructions on the following pages (for
additional lead placement options, see “Alternative Lead Placements” on
pg. 6-8).
For information on using disposable sensors, see “Disposable Resting ECG
Sensors” on pg. 6-5. For information on using reusable sensors, see
“Reusable ECG Sensors” on pg. 6-6.
NOTE:
Apply sensors and connect the patient cable and
lead wires before entering patient information this allows
the sensors to adhere, stabilize, and improve conductivity.
!
WARNING
False Readings. The lead status check does not detect lead reversals.
Always ensure the correct lead placement is used for the lead
configuration selected.
!
WARNING
Misdiagnosis. Dirty, compromised, or unprepared skin may cause
incorrect readings.
Before applying sensors, ensure the skin is clean, unbroken, and properly
prepared.
!
WARNING
Infection. Lead placement on compromised skin may cause infection.
Before applying sensors, ensure the skin is clean, unbroken, and properly
prepared.
!
Caution
Patient skin irritation. When applying sensors, skin preparation,
electrolyte solution, or electrode material may cause skin irritation.
Monitor the electrode site and, if irritation occurs, use alternative
electrolyte or electrodes.
Summary of Contents for Atria 3100
Page 8: ...viii Atria Electrocardiograph Operating Instructions 70 00082 01 A...
Page 22: ...2 Equipment Setup 2 8 Atria Electrocardiograph Operating Instructions 70 00082 01 A...
Page 34: ...3 User Interface 3 12 Atria Electrocardiograph Operating Instructions 70 00082 01 A...
Page 72: ...6 Patient Preparation 6 10 Atria Electrocardiograph Operating Instructions 70 00082 01 A...
Page 82: ...7 Acquiring ECG Reports 7 10 Atria Electrocardiograph Operating Instructions 70 00082 01 A...
Page 94: ...8 Patient Records 8 12 Atria Electrocardiograph Operating Instructions 70 00082 01 A...
Page 104: ...9 Sample Printouts 9 10 Atria Electrocardiograph Operating Instructions 70 00082 01 A...
Page 112: ...10 Troubleshooting 10 8 Atria Electrocardiograph Operating Instructions 70 00082 01 A...
Page 128: ...11 Maintenance and Service 11 16 Atria Electrocardiograph Operating Instructions 70 00082 01 A...
Page 134: ...4 Atria Electrocardiograph Operating Instructions 70 00082 01 A...