INSTRUCTIONAL MANUAL
2750L-P/4500L-P/6000L-P
August, 2001
Rev. K
-81-
3.8.8
PROGRAM EXAMPLES
Table 3-19 lists some typical CIIL strings. Each one illustrates the versatility in programming which is a feature of the
power source. The strings listed transmit the information with the fewest possible characters. Decimal points need not
be included when followed by zeros. Spaces may not be included anywhere in the string.
Table 3-19
SAMPLE PROGRAMMING STRINGS
PROGRAM STRING
All three phase to 115 volts and current limit
to 2 amperes and enable external sync
FNC ACS :CH00 SET CURL2 SET FREQ400 SET VOLT 115
SET SYNI <CR><LF>
Phase B to 100 volts and current limit to 2
amps
FNC ACS :CH02 SET CURL.2E1 SET
FREQ4E2 SET
VOLT100 <CR><LF>
To power-up state open relay
RST ACS :CH00 <CR><LF>
Phase B to 90 degrees
FNC ACS :CH02 SET PANG90 <CR><LF>
Freq. to 65.23 Hz
FNC ACS :CH00 SET FREQ65.23 <CR><LF>
All three phases to 2.5 amps average current
limit
FNC ACS :CH00 SET CURL2.5 <CR><LF>
Phase C current limit to 2 amps average
FNC ACS :CH03 SET CURL2 <CR><LF>
Confidence test
CNF <CRL><LF>
For status msg
STA <CR><LF>
Output relay closed
CLS :CH00 <CR><LF>
To measure output voltage only of Phase A
FNC ACS VOLT :CH01 <CR><LF>
Output relay opened
OPN :CH00 <CR><LF>
To measure Freq.
FNC ACS FREQ :CH00 <CR><LF>
FTH FREQ <CR><LF>
To measure output voltage of all phases
FNC ACS VOLT <CR><LF>
Query time of measurement of voltage
FNC ACS VOLT <CRl><LF>
INX VOLT <CR><LF>
Summary of Contents for 12000L-1P
Page 119: ...INSTRUCTIONAL MANUAL 2750L P 4500L P 6000L P August 2001 Rev K 115 Figure 6 1 MODULE LOCATION...
Page 138: ...INSTRUCTIONAL MANUAL 2750L P 4500L P 6000L P August 2001 Rev K 134 Figure 14 EMERGENCY VOLTAGE...
Page 148: ...INSTRUCTIONAL MANUAL 2750L P 4500L P 6000L P August 2001 Rev K 144 FIGURE 2...