INSTRUCTIONAL MANUAL
2750L-P/4500L-P/6000L-P
August, 2001
Rev. K
-58-
The following example will simultaneously ramp the Frequency from 400 to 500 Hz at a rate of 0.1 Hz per second and
the Amplitude from 5 volts in increments of .5 volts per step:
RNG270 AMP5 FRQ400 STP.1 DLY1 VAL500 STP.5
A GPIB Service Request will be generated at the end of a ramp if SRQ2 is included in the setup string.
A ramp operation can be stopped at anytime by the GPIB Group Execute Trigger.
3.7.11
TO PROGRAM A REGISTER (REG)
The REG header is used to load the register specified by the following numeric data with the preceding data. The
numeric value is from 0 to 15. The PRG header is identical to the REG header and is included to standardize other AC
power controllers.
The following example will load a ramp program that will step the voltage from 10 to 115 volts with 1 volt/.5 sec steps at
400 Hz into register 0.
FRQ400 AMP10 DLY.5 STP1 VAL115 REG0
3.7.12
TO RECALL A REGISTER (REC)
The REC header is used to recall previously loaded data from a register identified by the following register number (0 to
15).
The following example recalls and outputs the parameters stored in register 0 by an example in paragraph 3.7.11.
REC0
The following example recalls the parameters in register 0 and outputs the parameters after the IEEE-488 "GET"
message.
REC0 TRG
The following is an example of register linking. The voltage and frequency is maintained at 115 volts and 60 Hz for 5
seconds and then the program contained in register 0 is recalled and executed. The program is stored in register 1.
FRQ60 AMP115 DLY5 VAL115 REC0 REG1
The program is initiated by the following ASCII string:
REC1