INSTRUCTIONAL MANUAL
2750L-P/4500L-P/6000L-P
August, 2001
Rev. K
-43-
2.
Select the FRQ screen and specify the starting frequency of 360 Hz.
6
ENT 360
PRG
3.
Select the RMP A screen and specify the ramp parameters of the independent parameter, FRQ, of
DLY = .2 seconds, STP = .2 Hz and VAL = 440 Hz:
9
ENT A
0.2
PRG B
0.2 PRG C 440 PRG
4.
Select the RMP B screen and specify the ramp parameter of the dependent parameter, AMP, of STP
= .5 volts:
10
ENT B
0.5
PRG
5.
At this point the program may be executed by depressing the ENT key or stored in a register.
3.5.15 ERROR MESSAGES
Table 3-9 shows all of the possible error messages displayed on the front panel display. The cause of the error mes-
sage is also shown.
3.5.16
TO PROGRAM FREQUENCY, VOLTAGE, VOLTAGE RANGE AND CURRENT LIMIT DEFAULT
VALUES
The default values are the values that appear at power-up and after the GPIB Device Clear command.
To set any of the default values perform the following steps:
1.
Depress the MNU key several times until the first menu screen is displayed as illustrated below:
SNC = 01
*CLK = 02
*WVF = 03
RNG = 04
NOTE (*) May not be displayed. Depends on options
2.
Enter the key sequence: 9 5 9 ENT