INSTRUCTIONAL MANUAL
2750L-P/4500L-P/6000L-P
August, 2001
Rev. K
-146-
Table 1
NOTES: 1.
Definitions
T1 - Power Interrupt time.
T2 - Time it would take for the applied voltage to decay from V (nom) to zero volts >
T3 – Time it would take for the applied voltage to rise from zero to V (nom) volts.
V MIN – The minimum level (expressed as a percentage of V NOMINAL) to which the applied
voltage is permitted to decay.
2.
Tolerance to T1, T2 and V MIN = ±10%
Test
Condition
No.
1 2 3 4 5 6 7 8 9 10 11
12 13 14 15
T1
(MILLISECONDS)
2 10 25 50 75 100 200 1000 10 25 50 75 100 200 1000
T2
(MILLISECONDS)
<1 20* 20 20 20 20 20 20 50* 50* 50 50 50 50 50
T3
(MILLISECONDS)
<1 8 5 5 5 5 5 5 20 20 20
20 20 20 20
% OF V NOMINAL (V MIN)
0
50
15
10
5
0
0
0
20
50
0
15
5
0
0
•
VOLTAGE WILL NOT REACH ZERO IN THIS TEST CONDITION