ERROR MESSAGE
CAUSE
CRL FAULT
Indicate output current exceeds program current.
AMP (1) FAULT
Incorrect sense line connection. Overload on indicated output.
TEMP (1) FAULT
Amplifier overtemperature
CPU MEMORY FAULT
CPU failed self-test
DMA OVERFLOW
Remote message greater than 256 bytes.
EXT SYNC ERROR
No external sync input at System Interface connector. Signal is not between 45 and 2
or 5 KHz.
BUS LOCAL ERROR
Remote message sent while AC Power System is in local.
SYNTAX ERROR
Incorrect syntax received from IEEE-488 External Interface
AMP RANGE ERROR
Attempt to program AMP value greater than RNG value.
FRQ RANGE ERROR
Attempt to program FRQ less than 45 (2) or greater than 2 or 5 KHz.
PHZ RANGE ERROR
Attempt to program PHZ greater than ±999.9
CRL RANGE ERROR
Attempt to program CRL greater than 11.11 or 33.33.
RNG RANGE ERROR
Attempt to program RNG greater than 270.0 or other optional high voltage range.
RMPA RANGE ERROR
Attempt to program STP, DLY or VAL greater than the maximum.
DIV
ERROR
Consult
factory.
OVERFLOW ERROR
Consult factory.
_________________________________________________________________
(1) May be any combination of A, B or C.
(2) May be down to 17 Hz with reduced output voltage.
Table 3-9
FRONT PANEL DISPLAY ERROR MESSAGES