INSTRUCTIONAL MANUAL
2750L-P/4500L-P/6000L-P
August, 2001
Rev. K
-93-
To calibrate the measured power value send the following string:
CAL PWR (1) (Desired power value for phase (1)
NOTE: (1) May be blank or include an A, B or C extension.
4.3.6 REMOTE
OUTPUT
VOLTAGE CALIBRATION
The ASCII # symbol can be used with a remote calibration to greatly reduce the calibration time. The ASCII # is used in
place of a calibration coefficient to begin the CAL ramp to start at the existing CAL coefficient and ramp up or down for
a VAL argument of 255 or 0 respectively.
The following string will cause the CAL coefficient to ramp down with a delay of .2 seconds per step:
CAL # DLY.2 VAL0