MN260 B | 39
Service Manual
7.8.2.4 Apply cap (1137-SEQ) to Product Tank outlet fitting to
prevent ATF desaturation!
cap off
fitting
Figure 7-45. Cap
7.8.2.5 Apply 2 Cable Ties as shown on figure.
NOTE: Pre Loop Zip Ties.
Figure 7-46. Cable Ties
7.8.2.6 Attach the Exhaust Filter, Tube, Muffler Assembly and Tube, Muffler as shown in figure below.
NOTE: Ensure Arrow on Filter points in the correct direction. Install Cable Ties (6968-SEQ) as shown in figure below. Ensure the com-
pressor exhaust elbow tubing is through the bulkhead before/during assembly.
While applying Kapton tape, pull muffler tube to establish clearance. Make sure that tube, muffler, and muffler tube are laying
down on ATF surface.
NOTE: Pulling too hard on the Exhaust Tube may cause the tube to tear.
exhaust filter
(9221-SEQ)
tube
(9227-SEQ)
2x3” kapton tape
(2696-SEQ)
muffler assy
(SP9314-SEQ)
muffler tube
(9390-SEQ)
pull gently
here
ensure
clearance
Figure 7-47. Open SAROS