7.8.7.9 Add Cable Tie to hold wires away from the coil. Hand-
tighten loose as in Figure below.
cable tie
(6968-SEQ)
Figure 7-69. Cable Ties
7.8.7.10 Attach Top STRUT (9778-SEQ) with 5 screws as shown on figure below. Torque the screws to 6.0 in-lbs.
screws
(9912-1-SEQ)
STRUT
(9978-SEQ)
Figure 7-70. Top strut
7.8.8 POWER CONNECTOR REPLACEMENT (9766-SEQ)
7.8.8.1 After performing steps 6.7.1.1 thru 6.7.1.8 from “INLET CAP REMOVAL”
7.8.8.2 Remove the four screws (9203-2-SEQ) and remove the
Power Connector (9766-SEQ).
screws
power connector
(9766)
Figure 7-71. Remove screws
7.8.8.3 Install new Power Connector (9766-SEQ). Use Loctite
222 on the four mounting screws. Replace O-Ring (9916-3-
SEQ) if necessary.
Figure 7-72. Power Connector
7.8.8.4 Perform steps 7.8.10.4 – 7.8.10.11 from section “Inlet
Cap Installation”.