Service Manual
50 | MN260 B
7.8.10.5 Insert Power Harness wires from the Extended Top
Cap and tighten the screws to 4.4 in-lbs. Route AC power wires
through ferrite attached to board.
NOTE: Battery wire polarity indicated on the PCBA. Red is (+) and
Black is (-).
power
harness
Figure 7-81. Power Harness Wires
7.8.10.6 Insert Front Panel Flex cable flush with locking con-
nector. Apply 2” of Kapton Tape (2696-SEQ) on the Mounting
STRUT.
front panel
flex cable
securing tape
(2696-SEQ)
Figure 7-82. Flex cable
7.8.10.7 Connect the Ambient Air Thermistor and connect 9V
Battery Connector as shown.
9v battery
connector
Ambient Air
Thermistor
Figure 7-83. Ambient Air Thermistor
7.8.10.8 Attach the Oxygen Outlet Tube (6981-SEQ) to the Flow
Valve outlet port and secure with a cable tie.
NOTE: If flow valve is not installed, refer to Flow Valve Replace-
ment Procedure.
oxygen
tube
Figure 7-84. Oxygen Outlet Tube
Summary of Contents for SAROS 3000