Service Manual
22 | MN260 B
4.0 REPROGRAMMING/CALIBRATION
All reprogramming of the circuit boards shall be performed by CAIRE, Inc or by a factory authorized service center. Flow calibra-
tion is required whenever replacing an ATF, control board, compressor, or proportional valve. Flow calibration can be done at a
CAIRE authorized service center or in the field using item T-10560. See manual MN053 for details on performing flow calibration
and firmware updates in the field using item T-10560.
5.0 SHIPPING AND TRANSPORTING THE SAROS
When shipping the SAROS use original packaging if possible. Always remove the Battery from the SAROS before shipping.
If original packaging material is available repack the SAROS, Battery and power supplies in the designated packaging area.
If original packaging material or other CAIRE Inc. authorized shipping container is not available, contact CAIRE Inc. for replace-
ment shipping container.
6.0 STORING AND DISCARDING THE SAROS
WARNING: Do not expose the SAROS to water. The SAROS enclosure does not provide protection against the
harmful effects of liquid ingress. Electrical shock or damage to the unit may result.
STORING THE SAROS
Heat and humidity may degrade the performance or severely damage the SAROS. Store the device in a cool, dry protected area
away from high temperatures, moisture and humidity. Remove the Battery when storing the device.
Ensure that all recommended maintenance procedures in section 1.0 are performed while the device is in storage. It is especially
important that the 3-month requirement to run the device and fully drain the SAROS battery be performed while the unit is in
storage to ensure proper operation.
Discarding the SAROS
Battery Disposal: Your Battery is rechargeable and can be recycled. Always return it to an authorized service center or CAIRE
Inc. for proper disposal. You can also contact your local city or town offices for instructions on proper disposal of
the Battery.
SAROS: Local environmental laws may prohibit disposal of electrical and/or electronic equipment such as the SAROS and AC
Power Adapter. Contact the local city, town or country offices for instructions on proper disposal of electrical or electronic equip-
ment. Alternately, CAIRE Inc. may be contacted for disposal information at 1-800-482-2473.
Summary of Contents for SAROS 3000