Service Manual
20 | MN260 B
c) Complete this step using one of the methods below:
i) If using test equipment that automatically triggers a breath follow the manufacturer’s instructions for operation of the
test equipment. Record the average bolus reading.
ii) If manually triggering breaths, note and record the bolus size in mL of each breath. Repeat 5 times. Average the five
bolus measurements and record the average reading. You may need to “zero” the testing device between triggering a
simulated breath.
d) Adjust the SAROS to a setting of P96 (96mL).
e) Repeat steps B & C for the P96 testing.
3.2.3 Battery Run Time Test
1) Test battery run time using the following procedure:
a) Fully charge the battery via AC power by leaving it connected to AC power for a minimum of 1 hour. After 1 hour, verify the
power cartridge status indicator is fully illuminated and no longer displays a waterfall motion.
b) Disconnect the AC power. Power the SAROS on at C3 (3.0 LPM) and start a timer.
c) Allow the unit to continue to run until it shuts off.
d) Once the unit shuts off completely, stop the timer and record the battery duration.
e) Verify that the time recorded is >24 minutes
NOTE: The nominal capacity of a new battery is 30 minutes at setting C3. A battery is considered within its useful life if its duration is
80% of the nominal capacity of a new battery (24 minutes at C3).
3.0 ALERTS, ALARMS, AND TROUBLESHOOTING TABLE
The tables show the possible audible and visual alerts and alarms, their conditions and suggest appropriate troubleshooting
responses. The SAROS runs functional verification on alarm system as part of its operation. If you are unsure about any alerts or
alarm conditions, or responses please contact an authorized service technician or go to www.caireinc.com.
Alarm levels
Type
Level
Yellow LED
Audio Alarm
LCD
Description
Alarm
1
ON
1 beep 200ms only or repeats
every 20 seconds for 120 seconds
depending on the alarm
Displays Alarm
Category
Attention is
required
Status
0
OFF
OFF
NC (No Change)
All OK
Table 3-1. Alarm Levels
WARNING: AVAILABILITY OF A BACK-UP SOURCE OF OXYGEN IS RECOMMENDED IN CASE OF POWER OUTAGE OR A
DEVICE FAILURE. CONSULT YOUR PROVIDER FOR BACK-UP OXYGEN SYSTEM.
WARNING: DO NOT Ignore Alarms. System will attempt to produce oxygen under fault condition but may not be at
purity or flow.
Summary of Contents for SAROS 3000
Page 56: ...Service Manual 56 MN260 B NOTES...