Service Manual
40 | MN260 B
7.8.2.7 Install the ATF Product Pipe onto the Product Tank with Silicone tubing and secure with Cable ties. Install the other end
onto the ATF with the Product Tube and Clamp as shown in figure below.
0.5” silicone tube
(SP6981-1-SEQ)
cable tie
x2 (6968-SEQ)
ATF product
pipe (9342-SEQ)
clamp x2
(9369-SEQ)
ATF product
tube (9797-SEQ)
Figure 7-48. Open SAROS
7.8.3 RAIL INSTALLATION
7.8.3.1 Attach RIGHT STRUT (9780-SEQ) with five screws (9912-1-SEQ) with and torque to 6.0 in-lbs.
NOTE: Top bulkhead screw not shown.
Figure 7-49. Right Strut
7.8.3.2 Rotate assembly, route the ATF wires (stepper motor and
ATF thermistor wires) under the slot of the LEFT STRUT (9779-
SEQ).
NOTE: Pre-install cable ties in rail holes prior to installation.
Slot in
rail
Figure 7-50. Route ATF wires
7.8.3.3 Secure the STRUT (9779-SEQ) with five screws and
torque the screws to 6.0 in-lb.
ATF wires, verify
loop for proper
length
LEFT STRUT
(9779-SEQ)
Figure 7-51. Secure
Summary of Contents for SAROS 3000
Page 56: ...Service Manual 56 MN260 B NOTES...