MN260 B | 31
Service Manual
7.7.3.5 Loosen the Compressor Power screws and remove the wires.
5-Loosen
5-Loosen
1-Remove
2-Loosen
3-Disconnect
4-Remove
Figure 7-17. Battery Power Wires for Battery Interface
7.7.3.6 Disconnect Battery Communication Connector.
7.7.3.7 Disconnect the ATF Stepper Motor Connector from the
PCB.
Figure 7-18. Communication Connector for Battery Interface
ATF Stepper Motor Power, Fan Disconnect, ATF Thermistor
7.7.3.9 Cut the cable tie and remove the Flow Tube from the
Product Tank.
NOTE: It is recommended to cap this outlet using a piece of
tape during the repair process. Do not leave the plug open to
atmosphere longer than it takes to preform the repair process. If
this port is exposed longer than 24 hours, it must be completely
sealed to prevent damage to the ATF. If just exposed during repair,
tape is sufficient to prevent any damage.
Figure 7-19. ATF Thermistor Disconnect
7.7.3.10 Remove 2 screws on the Left STRUT (9779-SEQ).
7.7.3.11 Remove 2 screws on the Right STRUT (9780-SEQ).
Fan Connector
(red and black
wires)
ATF Thermistor
(blue wires)
ATF stepper
motor connector
(4 pin)
Battery communications
connector (5 pin)
• Cut Cable Tie
• Remove Tubing
• Cap Brass
Fitting
Summary of Contents for SAROS 3000