Interface Overview
88
©2020 CAE 905K710052 v.1.1
CAE
Air1
Operation Screen - PS Mode Backup Parameters
When the ventilator is in PS mode, additional backup control parameters are in place to increase the
breathing support when needed. The additional control parameters are Pi, RR and I:E.
If the patient is breathing on their own, these backup controls are inactive. If the patient becomes
apneic (not breathing) and the Tapnea threshold is surpassed, these backup control parameters are
activated. The active status is displayed onscreen on the control widgets.
Operation Screen - PS Mode
Backup Parameters showing Active Status
Note:
When the backup parameters are inactive, there is no onscreen indication.