Appendix B - Alarms
134
©2020 CAE 905K710052 v.1.1
CAE
Air1
Alarm Indications
When an alarm is triggered, audible and visible alerts signal the alarm condition and identify the
severity of the issue. All alarms can be silenced for two-minute intervals using the
Silence Alarms
button on the touchscreen. Once the triggering condition is resolved, the alarm is automatically reset
and silenced. The touchscreen is cleared of inactive banners by selecting the
Reset Alarms
button.
Audible Alarms
Ventilator alarm conditions are indicated by an audible alarm. High priority alarms will sound at the
loudest decibel setting and are not adjustable while medium and low priority alarms will sound at a
lower decibel level that can be set by the medical personnel.
Note:
The volume setting cannot be lowered below 25% of the maximum volume.
Visible Alarms
Ventilator alarm conditions are indicated by visible alarms in the form of an LED on the ventilator
case and an alarm banner on the touchscreen. The priority of the alarm is indicated by the colour as
well as by flashing or steady signals.
Alarm
Default
Range
Description
High Priority
85 dBA
85 dBA
High speed pulse-burst signal.
Maximum volume signal.
Volume cannot be altered.
Medium Priority
60 dBA
15-60 dBA
Medium speed pulse-burst signal.
Volume is adjustable.
Low Priority
60 dBA
15-60 dBA
Single pulse or two medium speed pulses
Volume is adjustable.
Alarm
Indication
Location
Description
High Priority
Red LED
Ventilator
Case
Flashing red omni-directional LED.
Red Banner
Touchscreen
Red banner indicating the affected sensor
and alert recorded in the Alarm History log.
Medium Priority
Yellow LED
Ventilator
Case
Flashing yellow omni-directional LED.
Yellow
Banner
Touchscreen
Yellow banner and alert recorded in the
Alarm History log.
Low Priority
Yellow LED
Ventilator
Case
Solid yellow omni-directional LED.
Yellow
Banner
Touchscreen
Yellow banner and alert recorded in the
Alarm History log.
Summary of Contents for Air1
Page 1: ...CAEAir1 Mechanical Ventilator User Guide...
Page 7: ...vi 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Page 9: ...Specifications 2 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Page 33: ...Safety 26 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Page 37: ...Introduction 30 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Page 45: ...Equipment Overview 38 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Page 131: ...Troubleshooting 126 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Page 153: ...Appendix F Applicable Standards 148 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...