©2020 CAE 905K710052 v.1.1
121
Care and Maintenance
Once the file is selected, update starts.
Uploading In Progress Screen
When the update is complete, the device will reboot. The page will indicate that the
browser may be closed.
Upload Complete
Note:
Software updates, service bulletins and a soft copy of this user guide are available on the
customer website at
www.caehealthcare.force.com/caeair1
.
Logs Collection
There are four types of logs which can be collected from a web browser:
• Alarms Log - records all information regarding all alarms triggered and cleared. Refer to
Appendix D - Ventilator Fault Logs
for details.
• Maintenance Log - a record of the self test result.
• System Log - records information intended only for CAE engineers.
• Platform Log - Any hardware malfunction will be shown in this log. Each error recorded in
the platform log has an ID associated with the entry and is defined in
Appendix D -
Ventilator Fault Logs. R
efer to this appendix for the lists of codes, error descriptions, and
appropriate actions for a technician to take for resolution.
Summary of Contents for Air1
Page 1: ...CAEAir1 Mechanical Ventilator User Guide...
Page 7: ...vi 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Page 9: ...Specifications 2 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Page 33: ...Safety 26 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Page 37: ...Introduction 30 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Page 45: ...Equipment Overview 38 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Page 131: ...Troubleshooting 126 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Page 153: ...Appendix F Applicable Standards 148 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...