Interface Overview
80
©2020 CAE 905K710052 v.1.1
CAE
Air1
Alarm Banners
Alarm conditions display as red or yellow banners at the top of the Operation screen. Red is used to
indicate a high priority alarm, while yellow is used for medium and low priority alarms.
If a large number of alarms are active at the same time, exceeding the banner display area, tap the
arrow to the right of the banners to expand the area to view all active alarms. Tap the arrow again to
return to the original view.
Alarm Banners - Expanded View
Clear Inactive Alarms
When alarms are active, they appear as red or yellow banners at the top of the Operation screen.
Low and medium priority banners are removed automatically when the alerting condition is
resolved. When high priority alarms are resolved, the alarm banner remains onscreen, but changes
colour from red to black to indicate that it is no longer active. The black banner remains displayed on
the screen until the
Clear Inactive Alarms
button is selected.
Clear Inactive Alarms Button
Silence Alarms
When alarms are active, an audible tone accompanies the visual indicators and the
Silence Alarms
button is enabled. To silence the alarms, tap the
Silence Alarms
button at the top of the Operation
screen. The audible alarms are paused for two minutes and then they will resume, unless the alerting
condition has been resolved. If a new alarm starts after the
Silence Alarms
button has been pressed,
all active alarms sounds will resume.
Silence Alarms Button
Note:
The visible alarms are not impacted by pressing the
Silence Alarms
button, only the audible
alarms are paused.
Summary of Contents for Air1
Page 1: ...CAEAir1 Mechanical Ventilator User Guide...
Page 7: ...vi 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Page 9: ...Specifications 2 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Page 33: ...Safety 26 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Page 37: ...Introduction 30 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Page 45: ...Equipment Overview 38 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Page 131: ...Troubleshooting 126 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Page 153: ...Appendix F Applicable Standards 148 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...