7.4
Cleaning the nebulizer
Materials Needed:
HPLC-grade 50:50 methanol: water solution
Sonication bath
u
Switch the On/Off master switch to Off.
u
Disconnect the liquid inlet line from the
nebulizer.
u
Disconnect the gas inlet from the nebulizer
u
Remove the knurled screws.
u
Remove the nebulizer from the drift tube.
u
Place the nebulizer in a beaker filled with 50:50
methanol: water solution.
u
Sonicate the nebulizer in an ultrasonic bath for
10 minutes.
u
If the nebulizer is still completely blocked,
connect a high-pressure air line to the nebulizer
inlet to help remove the blockage.
u
If the nebulizer cannot be cleaned, replace the
nebulizer.