Büchi Labortechnik AG
Operation | 6
Operation Manual Pure Chromatography Instruments
65/102
u
Follow the instructions on the display.
ð
Installing the cartridge. See Chapter 6.4.3 "Installing and removing a cartridge",
page 57
ð
Introducing the sample into the system. See Chapter 6.4.2 "Injecting a sample
into the flash system", page 55
6.6.4
Changings during a separation
NOTE
Parameters that can be edited are highlighted in green
Possibilities to edit the gradient during a separation:
Change the gradient. See Chapter "Editing the gradient in graphic mode", page 47
Hold the gradient. See Chapter 6.1.6 "Gradient panel", page 39
Auto gradient hold. See Chapter 6.1.6 "Gradient panel", page 39
6.6.5
Ending a Flash separation
NOTE
The separation extends automatically by 5 minutes if the baseline at the end of the
separation is not below the set threshold.
Precondition:
R
The display shows the dialog box
Separation End
.
u
According to the requirements extend the separation time by 5 minutes.
Precondition:
R
The display shows the dialog box
Run completed
.
u
Follow the instructions on the display.
ð
Purging the instrument with air. See Chapter 7.8 "Removing solvent from an
used cartridge", page 78
ð
Removing the cartridge. See Chapter 6.4.3 "Installing and removing a cartridge",
page 57
6.6.6
Shutting down the instrument
Navigation path
➔
File
Precondition:
R
The separation process has ended.
u
Purge the instrument with purging solvent. See Chapter 7.6 "Cleaning the
instrument", page 77
u
Navigate to the
[File]
menu via the navigation path.
u
Tap the action
[Shut down]
.
u
Confirm the secure question with
[Yes]
.
ð
The instrument is shutting down.